The Ezer EFC-2600 retinal camera is a fully automated non-mydriatic tabletop fundus camera built around 3D auto-tracking, autofocus and auto-capture, with a 45° field of view and a 12 MP sensor. It is the house-brand tabletop camera most US practices ask us about when they are adding retinal photography to a lane for the first time, or putting a second camera in a satellite office. Spec sheets for this unit are easy to find. What they do not tell you is how each of those numbers changes your room layout, your tech's workflow, and what you need to pin down in writing before you sign a purchase order.
This page is a single-model deep dive. If you are still deciding between form factors — tabletop non-mydriatic, mydriatic, or handheld — start with our fundus camera buying guide and come back here once the tabletop non-myd lane is the one you want. For the clinical rationale behind routine fundus photography, see our companion article on retinal imaging in eye care.
The published specifications
These figures are taken from the specification table on the live EFC-2600 product page. Every interpretation below is ours; every number below is theirs.
| Type | Digital non-mydriatic retinal camera |
| Photography modes | Color, digital red-free, anterior eye image |
| Image formats | JPEG, PNG, DICOM (optional) |
| Field of view | 45° |
| Minimum pupil size | 4.0 mm |
| Working distance | 25 mm |
| Focus adjustment range | −15D to +10D without compensation lens; −30D to −10D or +5D to +30D with compensation lens |
| Flash intensity | 10 levels, manually adjustable |
| Light source | Infrared LED (observation); white LED (flash) |
| Auto exposure | Yes |
| Image resolution | 12 MP |
| Eye fixation | Ten internal points |
| Alignment | Fully automatic 3D tracking |
| Chinrest movement | Motorized |
| Interfaces | USB 2.0, Ethernet, HDMI |
| Power supply | AC 100–240V, 50/60 Hz, auto-selected |
| Operating environment | 10°C to 35°C; 30% to 90% humidity, non-condensing |
| Dimensions (W × D × H) | 282 × 485 × 492 mm |
| Weight | 38 lb |
What the numbers actually mean for your lane
45° field of view, and what the mosaic function adds
45° is the standard single-capture field for tabletop non-mydriatic cameras, and it is what the EFC-2600 delivers per frame. The unit also offers an automatic mosaic function that stitches several fixation-target captures into one composite image — the product page lists the composite at up to 110°. Two things follow. First, a mosaic is several exposures, not one, so budget the extra seconds per patient when you are planning throughput. Second, a stitched composite is a different artifact from a true single-shot ultra-widefield image; if your workflow depends on widefield in a single capture, this is not that camera, and you should say so out loud during the demo rather than discover it later.
12 MP resolution and the imaging modes you will actually use
Twelve megapixels across a 45° field gives you room to zoom into a region without the image falling apart on screen — which matters more for patient education and for year-over-year comparison than it does for the initial capture. Beyond standard color, the camera does digital red-free imaging with adjustable gray scales, RGB channel filtering, and anterior segment photography. That last one is quietly useful: the same instrument that documents the posterior pole can also document an external finding, so one purchase covers two documentation jobs instead of one.
4.0 mm minimum pupil and 25 mm working distance
The 4.0 mm minimum pupil is the single most operationally important number on the sheet, and it is the one buyers most often skip. It defines the condition under which the camera can work without dilation. If a given patient's undilated pupil does not reach 4.0 mm in your dim exam room, that eye becomes a dilated capture — which is a scheduling and room-design fact, not a defect. Practices that get the most out of a non-mydriatic camera treat the darkened room as part of the equipment: a lane you can actually make dark, a few minutes of dark adaptation built into the flow, and a tech who knows to reduce ambient light rather than fight the instrument. The 25 mm working distance, meanwhile, is generous enough that positioning is comfortable for most patients and does not require the operator to crowd the patient's face.
Focus range and the compensation lens
Without a compensation lens the camera focuses from −15D to +10D, which covers the overwhelming majority of a general practice's patients. The compensation lens extends that to −30D to −10D or +5D to +30D. If you see a meaningful population of high myopes or aphakic patients, the compensation lens is not an optional nicety, it is part of your configuration — and it belongs on the quote from day one rather than as a follow-up order six weeks after install.
What "fully automatic 3D tracking" changes about your day
The automation stack here — 3D auto-alignment, autofocus, auto-exposure, motorized chinrest, ten internal fixation targets — is the real argument for this instrument, and it is a staffing argument more than an optical one. On a manually aligned camera, image quality tracks operator skill: your best tech gets good pictures and everyone else gets retakes. When alignment, focus and exposure are handled by the instrument, capture becomes a delegable task. That changes who can run the camera on a Saturday, how long training takes for a new hire, and how many retakes cost you lane time.
Two honest caveats. Automation reduces the skill floor for capture; it does not remove the need for someone who can recognize a bad image and take it again. And ten fixation targets are only useful if your staff is taught a repeatable capture protocol — otherwise you get inconsistent framing across visits, which quietly destroys the comparison value of the images you are paying to collect.
Room, table and installation planning
At 38 lb and 282 × 485 × 492 mm, the EFC-2600 is a permanent tabletop instrument, not something that moves between rooms. Plan for it accordingly:
- Table: it needs a stable powered instrument table rated for the weight, with enough depth for the 485 mm footprint plus patient-side clearance. If you are pairing it with another instrument on the same table, measure before you buy.
- Power: the auto-selecting 100–240V supply means no special electrical work, but you still want a dedicated, surge-protected outlet within reach of the table's final position.
- Light control: per the pupil requirement above, the room should be one you can genuinely darken. A camera in a bright hallway alcove will underperform its spec sheet forever.
- Environment: 10–35°C and 30–90% non-condensing humidity rules out unconditioned storage rooms and closets that swing with the seasons.
Image management, DICOM and getting pictures into the chart
The camera outputs JPEG and PNG natively, supports networking, and offers USB 2.0, Ethernet and HDMI interfaces. HDMI is more useful than it sounds — a second display facing the patient turns a capture into a conversation. Ethernet is how the images leave the instrument and reach a server or workstation rather than living on the device.
The one line to read twice: the specification table lists DICOM as optional. If your EHR or image-management pathway depends on DICOM, that is a line item to confirm in writing on your quote — not an assumption. This is the most common avoidable disappointment we see on imaging purchases across every brand: the integration everyone assumed was included turns out to be a separately licensed module, and it surfaces during go-live instead of during procurement.
Where the EFC-2600 fits — and where it does not
It fits a general optometric or ophthalmic practice adding tabletop retinal photography to a fixed lane; a growing practice putting a capable second camera in a satellite location; and any office where capture needs to be delegated to staff with varying experience rather than reserved for one skilled operator.
It is the wrong tool if you need imaging to travel. Mobile screening, nursing-home rounds, hospital ward consults and pediatric work all favor a handheld unit — our MiiS portable fundus camera is the counterpart here: also 45°, but 5 MP, 450 g, battery powered with 1.5–4.5 hours of operating time, a 3.5" touchscreen and Wi-Fi transfer. That is a genuinely different instrument for a genuinely different job, and plenty of practices end up owning both — the tabletop for the lane, the handheld for everything that cannot sit at a table. You can see the full lineup on our retinal cameras page.
Procurement checklist: confirm these before you sign
This is the part a spec sheet will never give you. Take this list into your quote conversation, for this camera or any other:
- Ask for the current spec sheet against your exact configuration. Third-party listings for this model circulate figures that do not match the manufacturer table — we have seen minimum pupil quoted as 3 mm elsewhere against the 4.0 mm on the official table, and mosaic coverage quoted at 100° in trade coverage against 110° on the product page. Buy against the document that will be attached to your purchase order, not against a search result.
- Confirm whether DICOM is included in your configuration or licensed separately, and confirm it against the specific EHR you run.
- Specify the compensation lens if you image high myopes or aphakic patients.
- Pin down training: who trains your staff, when, remote or on-site, and whether re-training after turnover is included.
- Get warranty terms and service response in writing — including who physically holds the spare parts and how long a repair keeps the camera out of the lane.
- Ask about the table. A camera quote without a table it can actually sit on is an incomplete quote.
Buying the EFC-2600 from US Ophthalmic
Ezer is our own house brand, which changes the ownership math in ways worth naming plainly. We are not passing your service call to a third party: US Ophthalmic runs an in-house technical department with factory-trained engineers, and our warehouse carries a stocked spare-parts inventory for immediate delivery — the difference between a camera that is down for a day and one that is down for a month. Every unit is backed by US warranty and service, and our 30-day money-back guarantee means the instrument has to earn its place in your lane before you are committed to it.
Configuration, availability and current pricing all move, so we quote per configuration rather than publishing a number that will be wrong next quarter. Request a quote on the EFC-2600 — tell us your lane, your patient mix and your EHR, and we will put together a quote that includes the accessories you will actually need, with availability confirmed against live stock.
Frequently asked questions
Does the Ezer EFC-2600 require pupil dilation?
It is a non-mydriatic camera, designed to image without dilation. The specification table lists a minimum pupil size of 4.0 mm, so imaging without dilation depends on the patient reaching that pupil size under dim room conditions. Practices that can properly darken the exam room get the most consistent results.
What field of view does the EFC-2600 have?
45° per single capture, which is the standard field for tabletop non-mydriatic cameras. Its automatic mosaic function combines multiple fixation-target captures into a composite image, listed on the product page at up to 110°.
Can staff without imaging experience operate it?
That is the design intent. Automatic 3D alignment, autofocus, auto-exposure and a motorized chinrest move most of the capture skill from the operator into the instrument, which makes the task delegable. You still want a trained protocol so framing stays consistent between visits.
Does the EFC-2600 integrate with an EHR?
It supports networking over Ethernet and outputs JPEG and PNG, with DICOM listed as an optional format. Because DICOM is optional rather than standard, confirm on your quote whether it is included in your configuration and validate it against your specific EHR before purchase.
How much space and what kind of table does it need?
The unit measures 282 × 485 × 492 mm and weighs 38 lb, so it needs a stable instrument table rated for the weight with room for the 485 mm depth plus patient clearance. It runs on standard AC 100–240V power, so no special electrical work is required.
Should I buy the EFC-2600 or a handheld camera?
Choose the EFC-2600 when imaging happens in a fixed lane and you want automation and image quality. Choose a handheld such as the MiiS portable when imaging has to travel to the patient — mobile screening, ward rounds, or patients who cannot be positioned at a table. Practices with both needs frequently buy one of each rather than compromising on a single unit.
What kind of warranty and service comes with it?
Units are covered by US warranty and supported by our in-house technical department and stocked spare-parts inventory, and are backed by a 30-day money-back guarantee. Exact warranty length and service terms are confirmed on your quote — always get them in writing.