Streak Retinoscopy plays a fundamental role in the examination process within the field of general and specialized optometry. This objective measurement of the refractive power of the eye can be highly valuable for both healthcare professionals and patients. Through a more detailed observation of the patient’s retinal health, physicians are able to properly interpret important findings. Both general and specialized ophthalmology practices will benefit from the superior performance and ease of use of the Ezer EZ-RET-2600. This 3.5-volt streak retinoscope with retinoscopy fixation cards allows physicians to quickly determine a patient’s retinal health using state-of-the-art precision and flexibility. Identify issues rapidly and efficiently, and assess retinal health and function.
The enhanced light emission makes refractions easier than ever, producing one of the sharpest, brightest, and cleanest mires available on the market. Discover unmatched performance in a compact design with practical, easy-to-use features.

Superior Quality
The EZ-RET-2600 features high-end multilayer optics, a single control ring for vergence, rotation, and comfort, as well as a metal handle for long-lasting performance. This retinoscope is also dustproof, making it maintenance-free and long-lasting.

With a streak retinoscope, the ophthalmologist can quickly determine the refractive properties of the patient’s undilated eye by modifying the width of the linear beam.
In the lowest position, a divergent beam is also available. The EZ-RET-2600 is equipped with halogen and LED illumination, and a bright 3.5-volt light for a thorough examination.

Includes 4 Fixation Cards
The genuine fixation card set magnetically attaches to the Ezer retinoscope head, allowing quick and easy card changes. The fixation retinoscopy cards offer age-appropriate symbols and words, based on the Pierce MEM card, for dynamic retinoscopy. These fixation cards facilitate the assessment of accommodation.

Can be part of the following kits:
Kit 1
Includes: EZ-OPH-3600, EZ-RET-2600, EZ-HND-2600, EZ-CHG-3600, EZ-OPH/RET-CASE

Kit 2
Includes: EZ-OPH-2600, EZ-RET-2600, EZ-HND-2600, EZ-CHG-3600, EZ-OPH/RET-CASE

Kit 3
Includes: EZ-OPH-2600, EZ-RET-2600, EZ-HND-700 (2)

Kit 4
Includes: EZ-OPH-3600, EZ-RET-2600, EZ-HND-700 (2)

Kit 5
Includes: EZ-OPH-2600 (Lamp Type: Halogen), EZ-RET-2600 (Lamp Type: Halogen), EZ-WTC-5200 DC
Kit 6
Includes: EZ-OPH-2600 (Lamp Type: LED), EZ-RET-2600 (Lamp Type: LED), EZ-WTC-5200 DC

Kit 7
Includes: EZ-OPH-3600 (Lamp Type: Halogen), EZ-RET-2600 (Lamp Type: Halogen), EZ-WTC-5200 DC
Kit 8
Includes: EZ-OPH-3600 (Lamp Type: LED), EZ-RET-2600 (Lamp Type: LED), EZ-WTC-5200 DC

Specifications
| Optical Specifications | |
|---|---|
| Rotation | ≥ 190° |
| Distance from the device at which the real image of the filament lamp is adjustable | ≤ 18 in / 450 mm |
| Distance from the device at which the virtual image of the filament lamp is adjustable | ≤ 18 in / 450 mm |
| Slit image length | ≥ 1.18 in / 30 mm |
| Slit image width | ≤ 0.06 in / 1.5 mm |
| Slit image linearity deviation at focus | ≤ 0.08 in / 2 mm |
| Rotational decentration of the slit image center | ≤ 0.4 in / 10 mm |
| All distances are measured from the light output of the device. When the focus is at 500 mm | |
| Electrical Specifications | |
| Bulb | 3.5 V 2.7 W |
| Operation Mode | Intermittent, on-time must not exceed 2 min with off-time no less than 5 min |
| Environmental Requirements | |
| Operation | |
| Ambient Temperature | +10 °C - +35 °C |
| Relative Humidity | 30% - 75% |
| Atmospheric Pressure | 700 hPa - 1060 hPa |
| Transport | Conventional Shipping |
| Storage | |
| Ambient Temperature | -40 °C - +55 °C |
| Relative Humidity | 10% - 90% |
| Atmospheric Pressure | 500 hPa - 1060 hPa |