MaxLight® 20D

$189.00

This item is currently sold out.
Design

Each lens incorporates state of the art aspheric computer generated design and manufacturing techniques.
OI-14 through STD-LR are made of a superior optical material, which cuts fatiguing weight by 40%.
They are guaranteed for 3 years against breakage, coating abrasion and defects in material and workmanship.
These eight lenses also receive an exclusive hydrophobic coating, which repels water, smudges and dust, making cleaning easier.
The 90D Standard is also available with a larger, 32mm holding ring. Product Code: OI-STD-LR
Lenses also available in colored mounts. Contact Ocular Instruments for further information.

Technique

Commonly known indirect ophthalmoscopy techniques using either the slit lamp or binocular indirect ophthalmoscope should be used.
The silver ring on the lens holder should be held toward the patient's eye during examination. It is important to recognize that this unidirectional design provides the best image quality possible.
Keep the lens centered on the patient's pupil.
Hold the lens far enough from the patient's eye so that the retinal image is the same diameter as the lens.
Keep the illumination source as dim as possible to minimize reflections and loss of image contrast.
Use the Ocular Lens Cleaning Cloth (OLCC) to keep lens clean and minimize glare from the lens surface.

Cleaning

Clean with alcohol wipe.

Disinfecting
Soak In: OR
GLUTARALDEHYDE BLEACH
2% or 3.4% aqueous solution
Temperature per manufacturer instructions
Minimum exposure time = 20 minutes
BLEACH
10% solution mixed at:
1 part bleach to 9 parts cool tepid water
Recommended exposure time = 10 minutes
Caution: To avoid damage to the lens, do not exceed recommended exposure time.
Then:
Rinse lens thoroughly to remove disinfection solution. 3 cycles of 1 minute, with cool or tepid water is recommended. Dry carefully and place in a dry storage case.
NOTE: This lens is known to be compatible with:Asepti-Wipe, Cavi-cide, Cidex, Cidex OPA, DisCideWipe, Enviro-cide, H2O2 -3% and OptiCide.

Sterilizing

AUTOCLAVE: No
STERRAD: No
STERIS SYSTEM 1: No
ETO: Yes, see ETO Parameters
ETO Parameters: Minimum Time 1 hour, Temperature 130°F (54°C), Aeration Time 12 hours

WARNING: Never Steam Autoclave or Boil listed lenses. Never soak in Alcohol, Acetone or Other Solvents.
DO NOT sterilize lens in its plastic or wooden storage case.
Features

Ocular Maxlight 20 Diopter indirect diagnostic/laser aspheric lens for B.I.O. high-resolution viewing. 2.97x image mag. 0.34x laser spot mag. 50ºstatic FOV, 47mm working distance, 48mm clear aperture, 39g lens weight.